Standard Calibrations (US)
Calibration | Fx,Fy | Fz | Tx,Ty | Tz | Fx,Fy | Fz | Tx,Ty | Tz |
US-2250-13000 | 2250 lbf | 5250 lbf | 13000 lbf-in | 13000 lbf-in | 3/8 lbf | 1 lbf | 3 3/4 lbf-in | 1 7/8 lbf-in |
US-4500-26000 | 4500 lbf | 10500 lbf | 26000 lbf-in | 26000 lbf-in | 3/4 lbf | 2 lbf | 7 1/2 lbf-in | 3 3/4 lbf-in |
US-9000-52000 | 9000 lbf | 21000 lbf | 52000 lbf-in | 52000 lbf-in | 1 1/2 lbf | 4 lbf | 15 lbf-in | 7 1/2 lbf-in |
| SENSING RANGES | RESOLUTION |
View Standard (US) Complex Loading Graph
Metric Calibrations (SI)
Calibration | Fx,Fy | Fz | Tx,Ty | Tz | Fx,Fy | Fz | Tx,Ty | Tz |
SI-10000-1500 | 10 kN | 22 kN | 1.5 kNm | 1.5 kNm | 1/640 kN | 1/240 kN | 3/8000 kNm | 3/16000 kNm |
SI-20000-3000 | 20 kN | 44 kN | 3 kNm | 3 kNm | 1/320 kN | 1/120 kN | 3/4000 kNm | 3/8000 kNm |
SI-40000-6000 | 40 kN | 88 kN | 6 kNm | 6 kNm | 1/160 kN | 1/60 kN | 3/2000 kNm | 3/4000 kNm |
| SENSING RANGES | RESOLUTION |
View Metric (SI) Complex Loading Graph
*The resolution is typical for most applications and can be improved with filtering.
**Resolutions quoted are the effective resolution after dropping four counts of noise.
***Applied loads must be within range in each of the six axes for the F/T sensor to measure correctly.
ATI's F/T calibrations are traceable to the National Institute of Standards and Technology (NIST). ATI certifies sensors are calibrated in accordance with applicable ATI procedures. These procedures are compliant with the ISO 9001 standard to ensure that products are within ATI specifications.
Contact your sales representative for accuracy information.
|