Standard Calibrations (US)
Calibration | Fx,Fy | Fz | Tx,Ty | Tz | Fx,Fy | Fz | Tx,Ty | Tz |
US-1.8-0.4 | 1.8 lbf | 3.15 lbf | 0.4 lbf-in | 0.4 lbf-in | 1/3400 lbf | 1/2720 lbf | 7/92800 lbf-in | 1/18560 lbf-in |
US-3.6-0.8 | 3.6 lbf | 6.3 lbf | 0.8 lbf-in | 0.8 lbf-in | 1/1700 lbf | 1/1360 lbf | 7/46400 lbf-in | 1/9280 lbf-in |
US-7.2-1.6 | 7.2 lbf | 12.6 lbf | 1.6 lbf-in | 1.6 lbf-in | 1/850 lbf | 1/680 lbf | 7/23200 lbf-in | 1/4640 lbf-in |
| SENSING RANGES | RESOLUTION |
View Standard (US) Complex Loading Graph
Metric Calibrations (SI)
Calibration | Fx,Fy | Fz | Tx,Ty | Tz | Fx,Fy | Fz | Tx,Ty | Tz |
SI-8-0.05 | 8 N | 14.1 N | 50 Nmm | 50 Nmm | 1/682 N | 1/682 N | 3/364 Nmm | 5/728 Nmm |
SI-16-0.1 | 16 N | 28.2 N | 100 Nmm | 100 Nmm | 1/341 N | 1/341 N | 3/182 Nmm | 5/364 Nmm |
SI-32-0.2 | 32 N | 56.4 N | 200 Nmm | 200 Nmm | 1/171 N | 1/171 N | 3/92 Nmm | 5/184 Nmm |
| SENSING RANGES | RESOLUTION |
View Metric (SI) Complex Loading Graph
*The resolution is typical for most applications and can be improved with filtering.
**Resolutions quoted are the effective resolution after dropping four counts of noise.
***Applied loads must be within range in each of the six axes for the F/T sensor to measure correctly.
ATI's F/T calibrations are traceable to the National Institute of Standards and Technology (NIST). ATI certifies sensors are calibrated in accordance with applicable ATI procedures. These procedures are compliant with the ISO 9001 standard to ensure that products are within ATI specifications.
Contact your sales representative for accuracy information.
|